0519-83866868
|
CH
Nickel titanium file series
Custom function file M3-RT retreatment file
产品介绍
◆ Product characteristics
The piecewise removal of the glue tips in the root canal can fix the glue tips on the root canal file, and it is not easy to push the glue tips to the root canal and prepare the root canal simultaneously
◆ Product highlights
Resolve secondary treatment
产品详情

Order of operation:

3009 (16mm)

Apply slight pressure in the direction of the root tip. Encounter obstacles, do not force, take the X tablet to find out the cause of the use of hand stainless steel file to eliminate obstacles. Check instrument threads frequently and clean debris during operation. Stop operation until there is no material debris between threads. If there is still filling material in the root canal, continue to use 2508 (18mm)

2508 (18mm)

Apply slight pressure in the direction of the root tip. Encounter obstacles, do not force, take X film to find out the reason; Use a hand stainless steel file to remove impediments. Check instrument threads frequently and clean debris during operation. Stop operation until there is no material debris between threads. If there is still filling material in the root canal, continue to use 2007 (22mm) to remove the step using a pre-bent 08,10 or 15 gauge C+ file, length :18mm

2007 (22mm)

Apply slight pressure in the direction of the root tip. Encounter obstacles, do not force, take X film to find out the reason; Use a hand stainless steel file to remove impediments. During operation, check the instrument thread frequently and identify the debris. Stop operation until there is no material debris between threads. Clean the filling material at the nearest 1mm root canal with a C+ file to prevent debris from being pushed out of the apical hole

有问题吗?
益锐医疗技术顾问为您服务!
0519-83866868
info@udg-dental.com