0519-83866868
|
CH
Nickel titanium file series
Custom function file M3-PATH 02 taper channel file
产品介绍
◆ Continuous high-speed continuous rotation
The debris is continuously discharged upward to prevent the debris from pushing out of the apical hole and forming a postoperative reaction
◆ Instead of using a hand K-file
The cleaning efficiency is fast, the postoperative reaction small root canal has good concentricity and high smoothness
◆ Product highlights
Root canal treatment, dredge the root canal
产品参数

 

specification

Length

Tip shape

Color identification

Section shape

0mm

1mm

3mm

7mm

11mm

15.5mm

Rotational speed

torque

1302

21mm

25mm

28mm

31mm

conical

purple

square

 

0.13

0.15

0.19

0.27

0.35

0.44

300

3.0

1602

21mm

25mm

28mm

31mm

conical

white

square

 

0.16

0.18

0.22

0.30

0.38

0.47

350

1.5

1902

21mm

25mm

28mm

31mm

conical

yellow

square

 

0.19

 

0.21

0.25

0.33

0.41

0.50

350

1.5

产品详情

Order of operation:

Prior to preparation, make sure to establish a standardized linear path to the pulp cavity to ensure that the root canal file reaches the root tip or the first bend in a straight line
The crown is fully pre-opened to ensure that the lubricant enters one third of the root tip smoothly with the file and maintains the feel
Always keep instruments well lubricated, especially the root tip one-third. EDTA lubricant is recommended.
Short lift to ensure that the instrument does not rest at any point in the root canal.
The instrument should not reach the working length at once, and should be washed every 4 mm or so
Ensure that each instrument is prepared to form a smooth, repeatable access channel to ensure that subsequent instruments smoothly "glide" forward.

有问题吗?